HPLC troubleshooter

187 0 1
                                    

*Broad peaks

old mobile phase,Low flow rate,Column overload,Void volume at inlet,too long or too wide Tubing.

*Ghost peaks

Contamination of injector or column,Retained compound from previous injection.

*Negative peaks

Refractive index of mobile phase higher than that of solute,great difference in composition between sample solvent and mobile phase,Mobile phase more absortive than sample components.

*Peak fronting

Column overload,Sample solvent incompatible with mobile phase,Void volume at inlet.

*Rounded peaks

Column overload,Wrong sample solvent.

*Base line noise

Air in mobile phase,detector cell or pump,incomplete mixing,leak,Bad quality mobile phase,detector Cell contamination,Weak detector lamp.

*Peak tailing

wrong column,Sample solvent incompatible with mobile phase,Void volume at inlet.